Level 1

Calibration reports are identified and indexed by the pipette’s serial number, manufacturer, and size. The next recommended service date is based on a semi-annual maintenance schedule, unless otherwise specified by the client.

Preventative Maintenance and Accuracy Verification for non-GMP/GLP Applications:

  • Complete disassembly (as applicable), and internal and external inspection and cleaning of instrument.
  • Replacement of Teflon seal, O-ring, or grease, as applicable, and as recommended by the manufacturer.
  • Replacement of broken or worn out parts as required.
  • Sticker affixed to pipette with information such as date of service and service provider, the serial number of the instrument, technician’s initials, date of calibration, and next calibration due by month, day and year. Specific details concerning the information to be presented on the calibration label are determined with the customer during the contract review process.
  • Calibration of instrument, gravimetrically at three levels, and to Precision Calibration Services’ SOP 125–Level 5 specifications, or to your unique calibration needs as defined by the contract review process.
  • Toll-Free Service Centre telephone number and easy-to-use Service Ordering via our website.
  • All Measurement and Test Equipment used for calibration is SI through NIST traceable (copies of all NIST traceable Calibration Certificates are available upon request).
  • As-Left inaccuracy and imprecision data is provided for each pipette based upon the number of samples and volumes agreed upon during the contract review process. Precision Calibration Services has a variety of validated templates designed to meet your verification needs.
  • Environmental conditions required for weight to volume calculation ascertained prior to each calibration event.

Quality Assurance Certificate:

Level 1 service provides a Quality Assurance Certificate, (attachment A-1) for the group of pipettes provided by the end user.